Home   >   Product Showcase   >   FlowPrep 060 Accessory Literature

FlowPrep 060 Accessory Literature

FlowPrep Accessory Literature


Request a Quote



The FlowPrep 060 applies both heat and a stream of inert gas to the sample. The heat causes contaminants to desorb from the surface and the stream of inert gas sweeps them out of the sample tube. It lets you choose the temperature, gas, and flow rate best suited for your sample material and application. Needle valves allow you to introduce the flowing gas slowly to prevent fluidization of samples.

 

 


 

 

Features

  • Temperatures up to 400 °C
  • Temperature accuracy of 5 °C
  • Gas flow rates up to 50 cm³/min
  • Temperature status display
  • Individual gas flow control valves
  • Accepts He, N2, Ar and other non-corrosive gases

Specifications

Capability

  • Method: Flowing gas
  • Capacity: 6 stations
  • Gases: He, N2, Ar and others
  • Flow Rate: Up to 50 cm³/min
  • Temperature range: Ambient to 400 °C

Physical

  • Height: 43 cm (16.9 in.)
  • Width: 36 cm (14.3 in.)
  • Depth: 30 cm (11.9 in.)
  • Weight: 8.5 kg (19 lbs)

Electrical

  • Voltage: 100, 120, 220, 240
  • VAC ±10%
  • Power: 200 VA (Maximum)
  • Frequency: 50/60 Hz