Load a Sample
See also: |
- Place the sample preparation tool on a flat surface.
- Insert a brass sample plug into the sample preparation tool.
- Place a paper filter on top of the brass sample plug.
- Insert the sample tube on top of the paper filter and press down until the sample tube is firmly seated around the brass plug.
- Insert the sample into the sample tube and weigh the sample tube assembly with the sample. Make a note of the mass.
- Place a paper filter on the open end of the sample tube.
- Insert a brass plug on top of the paper filter.
- Firmly press the paper filter and brass plug into the sample tube.
- For Inorganic analysis only, place the open end of the sample tool spacer onto the anvil. The spacer tool is used for the first compression only.
- Place the lower end of the sample tube onto the anvil.
- Install the sample tube collar and adjust the collar when the sample tube is in place.
- Install the safety shield over the anvil, sample tube, and piston assembly.