Performing a Calibration
Verification
Verify the AccuPyc III is operating as intended.
A
calibration verification is performed by analyzing the volume standard that
comes with the instrument. Start the verification on the screen and
follow the on-screen prompts.
Navigate to Settings > Chamber > Verify to begin a calibration check.
Select a volume standard. Nominal cell volume is either 10 cm3 or 100 cm3, depending on the AccuPyc III size in use.
Place the selected volume standard in the sample chamber.
Start the verification on the screen.
The measured volume should be within tolerance of the volume standard.
The
tolerance is given as:
Tolerance = (nominal cell volume * 0.02%) + (standard
volume * 0.02%)